Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Recalled by Biomet, Inc. Due to Biomet Trauma ("Biomet") has initiated a recall of...

Date: April 29, 2013
Company: Biomet, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Biomet, Inc. directly.

Affected Products

Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems. Product Usage: The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.

Quantity: 75

Why Was This Recalled?

Biomet Trauma ("Biomet") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021. The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Biomet, Inc.

Biomet, Inc. has 250 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report