Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization Recalled by Biomet, Inc. Due to These screws have a single lead (instead of...

Date: November 20, 2015
Company: Biomet, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Biomet, Inc. directly.

Affected Products

A.L.P.S. 4.0mm Cancellous Locking Screws intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones.

Quantity: 182, 965

Why Was This Recalled?

These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not lock into the plate and backs out, a revision surgery may be necessary.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (50 states)Not affected

About Biomet, Inc.

Biomet, Inc. has 250 total recalls tracked by RecallDetector.

Related Recalls