Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Biomet 3i Narrow Right Angle Square Driver Tip. Driver tips Recalled by Biomet 3i, LLC Due to RASQ3N, a square, abutment screw driver tip has...

Date: August 10, 2015
Company: Biomet 3i, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Biomet 3i, LLC directly.

Affected Products

Biomet 3i Narrow Right Angle Square Driver Tip. Driver tips are attached to a surgical drilling unit to carry components such as implant mounts, healing abutments and screws for placement and removal. For dental and restorative procedures.

Quantity: 103 pieces

Why Was This Recalled?

RASQ3N, a square, abutment screw driver tip has been labeled and distributed as RASH3N, a hexed, abutment screw driver tip. Additionally, the RASQ3N received the RASH3N's laser marking "RASH3N" during manufacturing.

Where Was This Sold?

This product was distributed to 29 states: AK, AZ, CA, CO, CT, FL, GA, ID, IL, IA, KY, LA, MD, MA, MI, MS, NE, NJ, NY, NC, OH, OR, PA, SD, TN, TX, UT, VA, WA

Affected (29 states)Not affected

About Biomet 3i, LLC

Biomet 3i, LLC has 222 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report