Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
API NIH (REF 10400). ZYM B reagent is used and Recalled by BioMerieux SA Due to bioMerieux identified a visual defect and activity issue...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact BioMerieux SA directly.
Affected Products
API NIH (REF 10400). ZYM B reagent is used and included inside of API NH (REF 10400) and API Listeria (10300). All of the API's are standardized systems combining several biochemical tests that enable group or species identification of most microorganisms.
Quantity: 33,853
Why Was This Recalled?
bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About BioMerieux SA
BioMerieux SA has 45 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report