Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

WaterLase iPlus Dental Laser System Recalled by Biolase Technology Inc Due to Biolase is recalling the WaterLase iPlus Dental Laser...

Date: June 12, 2014
Company: Biolase Technology Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Biolase Technology Inc directly.

Affected Products

WaterLase iPlus Dental Laser System, Part No. 7200854-XX Software Upgrade Kit, Part No. 6201463.

Quantity: 117 units

Why Was This Recalled?

Biolase is recalling the WaterLase iPlus Dental Laser System because of a discrepancy present in the user interface software (version 1.7). If an error occurs in the procedure screen, the air setting will change to 1% and OFF.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Biolase Technology Inc

Biolase Technology Inc has 1 total recall tracked by RecallDetector.