Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / Recalled by BioFire Diagnostics, LLC Due to Due to real-time stability study failures resulting in...

Date: June 7, 2021
Company: BioFire Diagnostics, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact BioFire Diagnostics, LLC directly.

Affected Products

FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)

Quantity: 4,462 kits (Globally)

Why Was This Recalled?

Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.

Where Was This Sold?

This product was distributed to 47 states: AL, AK, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, TN, TX, UT, VT, VA, WA, WV, WI, WY, DC

Affected (47 states)Not affected

About BioFire Diagnostics, LLC

BioFire Diagnostics, LLC has 40 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report