Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

FilmArray BCID Panel Recalled by BioFire Diagnostics, LLC Due to There is an increased risk of false positive...

Date: January 22, 2020
Company: BioFire Diagnostics, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact BioFire Diagnostics, LLC directly.

Affected Products

FilmArray BCID Panel, RFIT-ASY-0127, 6, Rx Only, CE, IVD, UDI: 00815381020093 used with the following BACT/ALERT Blood Culture Bottle Catalog No./Description: 410851/BACT/ALERT¿ FA Plus 410852/BACT/ALERT¿ FN Plus 410853/BACT/ALERT¿ PF Plus 259790/BACT/ALERT¿ SN Product Usage: is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.

Quantity: 5,020 kits

Why Was This Recalled?

There is an increased risk of false positive Proteus results when using the Film Array Blood Culture Identification Panel (BCID) in vitro diagnostic test of blood cultures used in conjunction with specific media types. False positive Proteus results from this test may lead to patients receiving improper antibiotic treatment until culture results are available (within 24-48 hours).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About BioFire Diagnostics, LLC

BioFire Diagnostics, LLC has 40 total recalls tracked by RecallDetector.

Related Recalls