Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BioFire BCID2 Panel Recalled by BioFire Diagnostics, LLC Due to The firm has identified an increased risk of...

Date: January 26, 2021
Company: BioFire Diagnostics, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact BioFire Diagnostics, LLC directly.

Affected Products

BioFire BCID2 Panel; IVD, Rx Only Part No: RFIT-ASY-0147 / UDI: 00815381020338 - Product Usage: intended for use with FilmArray systems.

Quantity: 5,220 kits

Why Was This Recalled?

The firm has identified an increased risk of false positive Pseudomonas aeruginosa when BCID/BCID2 Panels are used with specific BD BACTEC" blood culture vials.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About BioFire Diagnostics, LLC

BioFire Diagnostics, LLC has 40 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report