Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set) Recalled by BEMER INT. AG Due to There potential that electromagnetic fields generated in the...

Date: January 13, 2021
Company: BEMER INT. AG
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact BEMER INT. AG directly.

Affected Products

Bemer-Set/ Pro-Set for the following models: REF. 410201 (BEMER Pro-Set), REF. 420201 (B.BOX Professional), REF. 420100 (B.BOX CLASSIC), REF. 410100 (BEMER Classic-Set), REF. 420200 (B.BOX Professional), and REF. 410200 (BEMER Pro - Set)

Quantity: 197,724 units

Why Was This Recalled?

There potential that electromagnetic fields generated in the therapy system may cause technical defects in a patient using an insulin pumps or a malfunction of a pacemaker.

Where Was This Sold?

US