Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Belmont Rapid Infuser Recalled by Belmont Instrument Corporation Due to Leak from a saline bag caused saline to...

Date: September 25, 2012
Company: Belmont Instrument Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Belmont Instrument Corporation directly.

Affected Products

Belmont Rapid Infuser, a Fluid Management System, Model FMS2000.

Quantity: 2,913 units

Why Was This Recalled?

Leak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical flash at the power entry area.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Belmont Instrument Corporation

Belmont Instrument Corporation has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report