Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
REF: 386862 Recalled by Becton Dickinson Infusion Therapy Systems Inc. Due to There is the potential that the safety shield...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson Infusion Therapy Systems Inc. directly.
Affected Products
REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO
Quantity: 98,000 units
Why Was This Recalled?
There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Becton Dickinson Infusion Therapy Systems Inc.
Becton Dickinson Infusion Therapy Systems Inc. has 4 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report