Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System Recalled by Becton Dickinson & Company Due to Potential development of cracks or breaks in the...

Date: June 19, 2020
Company: Becton Dickinson & Company
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Company directly.

Affected Products

PowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers AN132075 (UDI 00801741118968), AN142010 (UDI 00801741119231), AN142010G (UDI 00801741119323), AN142075 (UDI 00801741119224)

Quantity: 5,296,132 total

Why Was This Recalled?

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

Where Was This Sold?

Worldwide Distribution

About Becton Dickinson & Company

Becton Dickinson & Company has 169 total recalls tracked by RecallDetector.

Related Recalls