Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 Recalled by Becton Dickinson & Company Due to A portion of this lot of blood collection...

Date: January 16, 2017
Company: Becton Dickinson & Company
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Company directly.

Affected Products

BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 ml x 13 x 75 mm) Catalog Number: 369714 Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations

Quantity: 1,488,800 tubes

Why Was This Recalled?

A portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant.

Where Was This Sold?

This product was distributed nationwide across all 50 states.