Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use Recalled by Becton Dickinson & Company Due to BD identified that several lots did not receive...

Date: September 16, 2016
Company: Becton Dickinson & Company
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Company directly.

Affected Products

BD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection

Quantity: 648,000 units

Why Was This Recalled?

BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required of 10-6.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Becton Dickinson & Company

Becton Dickinson & Company has 169 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report