Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PrepStain Tecan US-II. For use in the screening and detection Recalled by Becton Dickinson & Co. Due to PrepStain system rack setting was incorrectly entered on...

Date: July 8, 2013
Company: Becton Dickinson & Co.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Co. directly.

Affected Products

PrepStain Tecan US-II. For use in the screening and detection of cervical cancer.

Quantity: 890 instruments

Why Was This Recalled?

PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Becton Dickinson & Co.

Becton Dickinson & Co. has 183 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report