Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended Recalled by Becton Dickinson & Co. Due to BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration...

Date: March 24, 2021
Company: Becton Dickinson & Co.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Co. directly.

Affected Products

BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal specimens Catalog Number: 444213

Quantity: 21,529 Kits

Why Was This Recalled?

BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Becton Dickinson & Co.