Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD Synapsys Informatics Solution Recalled by Becton Dickinson & Co. Due to There is a risk of mis-association of patient...

Date: June 8, 2023
Company: Becton Dickinson & Co.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Co. directly.

Affected Products

BD Synapsys Informatics Solution, Catalog Number 444150

Quantity: 102 systems

Why Was This Recalled?

There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.

Where Was This Sold?

This product was distributed to 15 states: CA, CO, IL, IA, MI, MN, MO, NY, OH, OK, SC, TN, TX, UT, WI

Affected (15 states)Not affected

About Becton Dickinson & Co.

Becton Dickinson & Co. has 183 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report