Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD Max Instrument Recalled by Becton Dickinson & Co. Due to A trend for false positive Candida glabrata results...

Date: September 24, 2018
Company: Becton Dickinson & Co.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Co. directly.

Affected Products

BD Max Instrument, Catalog Number 441916 441927 (refurbished instrument) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Quantity: 49

Why Was This Recalled?

A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Becton Dickinson & Co.

Becton Dickinson & Co. has 183 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report