Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD BBL Sensi Disc Ertapenem- 10 ug Recalled by Becton Dickinson & Co. Due to There is a possibility of reproducibility, accuracy, and/or...

Date: January 8, 2024
Company: Becton Dickinson & Co.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton Dickinson & Co. directly.

Affected Products

BD BBL Sensi Disc Ertapenem- 10 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 232174 and 232175¿

Quantity: 2,363,168 total units

Why Was This Recalled?

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Becton Dickinson & Co.

Becton Dickinson & Co. has 183 total recalls tracked by RecallDetector.