Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995) Recalled by Becton, Dickinson and Company, BD Biosciences Due to Due to partial label detachment on tubes. This...

Date: October 3, 2022
Company: Becton, Dickinson and Company, BD Biosciences
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Becton, Dickinson and Company, BD Biosciences directly.

Affected Products

BD Multitest 6-Color TBNK with optional BD Trucount Tubes (Cat. No. 662995), for use with BD FACSLyric, BD FACSCanto II, and BD FACSCanto flow cytometers to determine the percentages and absolute counts of human lymphocyte subsets in peripheral whole blood for immunophenotyping.

Quantity: 1,457 tubes (corrected on 11/04/2022)

Why Was This Recalled?

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant.

Where Was This Sold?

This product was distributed to 43 states: AL, AZ, AR, CA, CO, CT, DE, FL, GA, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, UT, VT, VA, WA, WI, DC

Affected (43 states)Not affected

About Becton, Dickinson and Company, BD Biosciences

Becton, Dickinson and Company, BD Biosciences has 35 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report