Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
BD FACSLyric Flow Cytometer Recalled by Becton, Dickinson and Company, BD Biosciences Due to Flow cytometers, when used with specific tubes may...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Becton, Dickinson and Company, BD Biosciences directly.
Affected Products
BD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD Trucount Tubes
Quantity: 22
Why Was This Recalled?
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Where Was This Sold?
This product was distributed to 26 states: CA, CO, FL, GA, HI, IL, IN, IA, KS, MD, MA, MI, MN, MO, NE, NJ, NY, OH, OK, PA, RI, TN, TX, VT, VA, WA
About Becton, Dickinson and Company, BD Biosciences
Becton, Dickinson and Company, BD Biosciences has 35 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report