Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SYNCHRON Systems Cholesterol (CHOL) Reagent Recalled by Beckman Coulter Inc. Due to Cholesterol Reagent may fail with Out of Calibration...

Date: November 10, 2021
Company: Beckman Coulter Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Quantity: 1433

Why Was This Recalled?

Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected