Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical Recalled by Beckman Coulter Inc. Due to Beckman Coulter has identified that due to a...

Date: August 3, 2017
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.

Quantity: 27 units

Why Was This Recalled?

Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.

Where Was This Sold?

China France Italy Kuwait Spain Turkey United States

About Beckman Coulter Inc.

Beckman Coulter Inc. has 241 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report