Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Microalbumin Reagent Recalled by Beckman Coulter Inc. Due to Beckman Coulter has confirmed that MA Reagent Lot...

Date: April 24, 2017
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON¿ Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine

Quantity: 3,158 units total (1,957 units in US)

Why Was This Recalled?

Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Beckman Coulter Inc.

Beckman Coulter Inc. has 241 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report