Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

iQ200 Series Urine Microscopy Analyzer- All part numbers Recalled by Beckman Coulter Inc. Due to A defect in the SATA power adapter cable...

Date: March 30, 2020
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

iQ200 Series Urine Microscopy Analyzer- All part numbers

Quantity: 3576 units

Why Was This Recalled?

A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Beckman Coulter Inc.

Beckman Coulter Inc. has 241 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report