Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Dxl 9000 Access Immunoassay Analyzer Recalled by Beckman Coulter, Inc. Due to The analyzers were built with nonconforming USB cables...

Date: June 4, 2024
Company: Beckman Coulter, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter, Inc. directly.

Affected Products

Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.

Quantity: 143 devices

Why Was This Recalled?

The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.