Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Discrete photometric chemistry analyzer for clinical use. Recalled by Beckman Coulter, Inc. Due to Beckman Coulter, Inc. has identified a rise in...

Date: June 14, 2024
Company: Beckman Coulter, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter, Inc. directly.

Affected Products

Discrete photometric chemistry analyzer for clinical use.

Quantity: 59 units

Why Was This Recalled?

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Beckman Coulter, Inc.

Beckman Coulter, Inc. has 79 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report