Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CellMek SPS Sample Preparation System Recalled by Beckman Coulter, Inc. Due to Sample preparation system designed to automate staining, lysing,...

Date: December 23, 2025
Company: Beckman Coulter, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter, Inc. directly.

Affected Products

CellMek SPS Sample Preparation System, REF: C44603, with software

Quantity: 50

Why Was This Recalled?

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected