Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Beckman Coulter Synchron AQUA CAL Level 1 Recalled by Beckman Coulter Inc Due to The firm identified a negative shift of 5mmol/L...

Date: May 15, 2017
Company: Beckman Coulter Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc directly.

Affected Products

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.

Quantity: 4444 AQUA CAL Level 1 kits

Why Was This Recalled?

The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.

Where Was This Sold?

This product was distributed nationwide across all 50 states.