Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Beckman Coulter Power Express Sample Processing System AU5800XL connection unit Recalled by Beckman Coulter Inc. Due to Potential exposure to biohazard. Software design problem causes...

Date: May 17, 2019
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.

Quantity: 291 units

Why Was This Recalled?

Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.

Where Was This Sold?

This product was distributed to 23 states: AL, CA, CO, DE, FL, GA, IL, IN, LA, MD, MA, MI, NJ, NY, NC, ND, OH, OR, PA, SD, TN, TX, WA

Affected (23 states)Not affected

About Beckman Coulter Inc.

Beckman Coulter Inc. has 241 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report