Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Beckman Coulter MicroScan PROMPT Inoculation System-D Recalled by Beckman Coulter Inc. Due to Due to customer reports it has been determined...

Date: December 13, 2019
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial susceptibility test.

Quantity: 3,472 units

Why Was This Recalled?

Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Beckman Coulter Inc.

Beckman Coulter Inc. has 241 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report