Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Beckman Coulter Lipase Recalled by Beckman Coulter Inc. Due to NAPQI, a metabolite of Acetaminophen, causes negative interference...

Date: August 24, 2018
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

Beckman Coulter Lipase, REF OSR6130 and OSR6230.

Quantity: 20,926 units

Why Was This Recalled?

NAPQI, a metabolite of Acetaminophen, causes negative interference with the recalled product assays at levels present in the serum/plasma of overdose patients.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Beckman Coulter Inc.

Beckman Coulter Inc. has 241 total recalls tracked by RecallDetector.

Related Recalls