Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated Recalled by Beckman Coulter Inc. Due to Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400...

Date: April 26, 2013
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum, urine, and other body fluids, including whole blood.

Quantity: 108 units

Why Was This Recalled?

Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman Coulter has received complaints that the joint connecting the tubing appears to disintegrate over time, causing a leak.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Beckman Coulter Inc.

Beckman Coulter Inc. has 241 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report