Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Access Thyroglobulin Calibrators Recalled by Beckman Coulter, Inc. Due to It has been determined through an internal investigation...

Date: July 6, 2023
Company: Beckman Coulter, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter, Inc. directly.

Affected Products

Access Thyroglobulin Calibrators

Quantity: 5980 units

Why Was This Recalled?

It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.

Where Was This Sold?

worldwide except US, Canada, and China

About Beckman Coulter, Inc.

Beckman Coulter, Inc. has 79 total recalls tracked by RecallDetector.