Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Access Thyroglobulin Assay Recalled by Beckman Coulter Inc. Due to Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent...

Date: July 28, 2015
Company: Beckman Coulter Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter Inc. directly.

Affected Products

Access Thyroglobulin Assay; 100 determinations, 2 packs, 50 tests/pack. For the quantitative determination of thyroglobulin levels in human serum and plasma using the Access Immunoassay Systems.

Quantity: 3,035 (U.S.) and 7,123 (International)

Why Was This Recalled?

Beckman Coulter, Inc. is recalling Access Thyroglobulin reagent packs because some may exhibit greater than 10% imprecision at concentrations <10 ng/mL.

Where Was This Sold?

This product was distributed nationwide across all 50 states.