Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD FACSCalibur Recalled by BD Biosciences, Systems & Reagents Due to The circuit breakers on some BD FACSCalibur instruments...

Date: March 6, 2015
Company: BD Biosciences, Systems & Reagents
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact BD Biosciences, Systems & Reagents directly.

Affected Products

BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

Quantity: 72 devices

Why Was This Recalled?

The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.

Where Was This Sold?

This product was distributed to 3 states: NJ, OK, TX