Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for Recalled by Bayer Medical Care, Inc. Due to The torque wrench used on the ceiling column...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Bayer Medical Care, Inc. directly.
Affected Products
Overhead Counterpoise Systems (OCS) OCS CONF an accessory device for use with Bayer injector systems as a support for an injector head. , SHORT CEIL MNT, STD HORZ ARM, Catalog # OCS 115AS, Material Number 59880288
Quantity: 1
Why Was This Recalled?
The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.
Where Was This Sold?
This product was distributed to 3 states: OH, TX, WV
About Bayer Medical Care, Inc.
Bayer Medical Care, Inc. has 8 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report