Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Welch Allyn CP150 Electrocardiograph with Spirometry Option Recalled by Baxter Healthcare Corporation Due to During manufacturing, the step to install an Electromagnetic...

Date: June 3, 2024
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

Quantity: 1 unit

Why Was This Recalled?

During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.

Where Was This Sold?

Distribution to Illinois.

About Baxter Healthcare Corporation

Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report