Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Oral Probe (Product code 02893-000) Recalled by Baxter Healthcare Corporation Due to Affected oral/axillary probes were inadvertently programmed with the...

Date: September 17, 2025
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers

Quantity: 5209

Why Was This Recalled?

Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Baxter Healthcare Corporation

Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report