Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Hillrom Q-Link 13 component Recalled by Baxter Healthcare Corporation Due to The Q-link 13 could allow for an improper...

Date: May 30, 2025
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2010004; 2) Viking L Mobile lift - Product Code 2040044; 3) Viking XL Mobile lift - Product Code 2040043; 4) Viking M Mobile lift - Product Code 2040045A; 5) Viking S Mobile Lift - Product Code 2040006; 6) Viking XS Mobile Lift - Product Code 2040007; 7) LikoLight Mobile lift - Product Code 2030001. .

Quantity: 9899 units

Why Was This Recalled?

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Where Was This Sold?

Worldwide

About Baxter Healthcare Corporation

Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report