Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Hill-Rom Progressa Bed System Recalled by Baxter Healthcare Corporation Due to The compression links on impacted Progressa beds, listed...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.
Affected Products
Hill-Rom Progressa Bed System
Quantity: 248 beds (Updated)
Why Was This Recalled?
The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur with a patient weight of less than 500 lbs. depending on the patient 's position on the bed.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Baxter Healthcare Corporation
Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report