Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

GEM FLOW COUPLER Monitor Recalled by Baxter Healthcare Corporation Due to An issue was identified where the LCD monitor...

Date: June 19, 2023
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

GEM FLOW COUPLER Monitor, PN 5156-00000-011

Quantity: 122 units

Why Was This Recalled?

An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.

Where Was This Sold?

US

About Baxter Healthcare Corporation

Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report