Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

FlowCOUPLER. An implantable device that is used to detect blood Recalled by Baxter Healthcare Corporation Due to Baxter Healthcare is recalling the FlowCOUPLER Device because...

Date: July 11, 2014
Company: Baxter Healthcare Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

FlowCOUPLER. An implantable device that is used to detect blood flow in vessels. Used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Baxter Healthcare.

Quantity: 1,315 units

Why Was This Recalled?

Baxter Healthcare is recalling the FlowCOUPLER Device because there is a possibility that some units may not maintain electrical continuity during handling intra-operatively and post-operatively, resulting in a loss of Doppler signal. This may result in the inability to monitor anastomotic patency, with the possibility that blood flow may not be appropriately monitored.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected