Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Baxter Operating Table TruSystem 7000 (MBW) Recalled by Baxter Healthcare Corporation Due to The firm received customer reports stating that the...

Date: September 19, 2024
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

Baxter Operating Table TruSystem 7000 (MBW), Product code REF 1841048

Quantity: N/A

Why Was This Recalled?

The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.

Where Was This Sold?

Worldwide distribution.

About Baxter Healthcare Corporation

Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report