Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Baxter NovumIQ Syringe INFUSION SYSTEM Recalled by Baxter Healthcare Corporation Due to Baxter Healthcare Corporation is issuing an Urgent Medical...

Date: March 3, 2025
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

Baxter NovumIQ Syringe INFUSION SYSTEM , REF 40800BAXUS

Quantity: 149 units

Why Was This Recalled?

Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Novum IQ syringe pumps listed below due to an incorrectly installed gasket. A gasket failure could lead to potential fluid ingress due to cleaning, or as a result of an IV fluid spill.

Where Was This Sold?

This product was distributed nationwide across all 50 states.