Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Allen Advance Table Recalled by Baxter Healthcare Corporation Due to the adhesive, applied to screws and bolts as...

Date: September 20, 2022
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

Allen Advance Table, RSP-ADVANCE TABLE: a) REF A-71100-US, #107003, b) A-71101-EU, #107003. operating room table

Quantity: 2 units

Why Was This Recalled?

the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.

Where Was This Sold?

This product was distributed to 4 states: MI, TX, WA, WI

Affected (4 states)Not affected

About Baxter Healthcare Corporation

Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report