Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope Recalled by Baxter Healthcare Corporation Due to The replacement rechargeable batteries inserted into the Welch...

Date: May 9, 2025
Company: Baxter Healthcare Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corporation directly.

Affected Products

2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600

Quantity: 277 units

Why Was This Recalled?

The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Baxter Healthcare Corporation

Baxter Healthcare Corporation has 401 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report