Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Four Lead TUR irrigation Set Recalled by Baxter Healthcare Corp. Due to A firm's internal investigation led to the discovery...

Date: May 15, 2013
Company: Baxter Healthcare Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corp. directly.

Affected Products

Four Lead TUR irrigation Set, for use with UROMATIC Plastic Containers, Product Code 2C4013 - Approximate Length 85" (2.2. m) Usage: Baxter irrigation sets are intended for the delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cytoscopy, transurethral resection and arthroscopic procedures. the sets are intended for use with Baxter irrigating solutions in Uromatic or Arthromatic plastic containers.

Quantity: 356,728 units

Why Was This Recalled?

A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinholes are not likely to be detected by the end user prior to use and could compromise the sterile barrier properties of the packaging.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Baxter Healthcare Corp.

Baxter Healthcare Corp. has 126 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report