Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Baxter's elastomeric infusion pumps are non-electronic Recalled by Baxter Healthcare Corp. Due to There are reported incidents of leakage on the...

Date: July 31, 2009
Company: Baxter Healthcare Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxter Healthcare Corp. directly.

Affected Products

Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication therapies requiring continuous infusion. The infusion pumps can be utilized for slow, continuous delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), and subcutaneous or epidural infusion of medications directly into an intraoperative site or subcutaneously for post-operative pain management.

Quantity: Code 2C1008KP: 56,400 units; Code 2C1009KP: 180,004 units; Code 2C1063KP: 154,937 units; Code 2C1087KP: 43,521 units; Code 2C1156KP: 7,908 units

Why Was This Recalled?

There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occurred after pumps are filled, primed, re-capped (using the blue winged Luer cap) and stored for a period of time, typically overnight. In addition, the reported instances of leakage

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Baxter Healthcare Corp.

Baxter Healthcare Corp. has 126 total recalls tracked by RecallDetector.