Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ExactaMix 1200 and 2400 Compounders. Used as an automated pumping Recalled by Baxa Corporation Due to Baxter Healthcare is recalling certain ExactaMix automated compounding...

Date: June 4, 2013
Company: Baxa Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Baxa Corporation directly.

Affected Products

ExactaMix 1200 and 2400 Compounders. Used as an automated pumping system that compounds multiple sterile ingredients into a finished solution.

Quantity: 753 units

Why Was This Recalled?

Baxter Healthcare is recalling certain ExactaMix automated compounding devices due to the complementary metal oxide semi-conductor battery failure resulting in resetting the internal clock of the device to midnight December 31, 2010.

Where Was This Sold?

This product was distributed nationwide across all 50 states.